Results

The synthetic ECG (Fig. 5) illustrates the modulation of the QRS-complex due to RSA. Observational uncertainty is incorporated by adding normally distributed measurement errors with mean zero and standard deviation 0.025 mV (Fig. 6a), yielding a similar signal to a segment of real ECG from a normal human (Fig. 6b). In order to illustrate the dynamics of the RR-intervals obtained from this synthetic ECG, peak detection was used to identify the times of the R-peaks. In the noise-free case, a simple algorithm which looks for local maxima within a small window is sufficient. For ECGs with noise and artefacts it may be necessary to use more complicated methods [2,3]. A comparison between the continuous process with power spectrum $S(f)$ given by (3) and the piecewise constant reconstruction of the RR-process obtained from the R-peak detection (Fig. 7) illustrates the measurement errors that arise when computing heart rate variability statistics from RR-intervals. The RR-intervals (Fig. 8a) and corresponding instantaneous heart rate (Fig. 8b) in units of beats per minute (bpm) for a mean of 60 bpm and standard deviation of 5 bpm display variability due to both RSA and Mayer waves. A spectral estimation technique for unevenly sampled time series, the Lomb periodogram [15,16], was used to calculate the power spectrum (Fig. 8c) from the RR tachogram, derived from 5 minutes of data as recommended by [7,10]. Despite the loss of information in going from the continuous process to the piecewise constant reconstruction, a comparison between Fig. 4 and Fig. 8c illustrates that it is still possible to obtain a reasonable estimate of the power spectrum.
Figure 5: ECG generated by dynamical model: (a) 10 seconds and (b) 50 seconds.
\begin{figure}
\centerline{\psfig{file=ecgsynthetic.eps,width=7.75cm}}
\end{figure}
Figure 6: Comparison between (a) synthetic ECG with additive normally distributed measurement errors and (b) real ECG signal from a normal human.
\begin{figure}
\centerline{\psfig{file=ecgcomparison.eps,width=7.75cm}}
\end{figure}
Figure 7: Reconstruction of RR-process from R-peak detection: the underlying RR-process generated using (3) (black line) and the RR-interval time series obtained using R-peak detection of the synthetic ECG (grey line).
\begin{figure}
\centerline{\psfig{file=rrinout.eps,width=7.75cm}}
\end{figure}
Figure 8: Analysis of RR-intervals from R-peak detection of the ECG signal generated by the dynamical model (1) with mean heart rate 60 bpm and standard deviation 5 bpm: (a) RR-intervals, (b) instantaneous heart rate and (c) power spectrum of the RR-intervals. Note the two active frequencies belonging to RSA (0.25 Hz) and Mayer waves (0.1 Hz).
\begin{figure}
\centerline{\psfig{file=rrsynthetic.eps,width=7.75cm}}
\end{figure}
An increase in the RR-interval implies that the trajectory has more time to get pushed into the peak and trough given by the R and S events. This is reflected by the strong correlation between the RR-intervals and the RS-amplitude as shown in Fig. 9. A technique for deriving a measure of the rate of respiration from the ECG has been proposed [5,6]. This ECG-derived respiratory signal (EDR) is of clinical use in situations where the ECG, but not respiration, is recorded. The synthetic ECG provides a means of testing the robustness of such techniques against noise and the effects of different sampling frequencies.
Figure 9: RS-amplitudes versus RR-intervals for the synthetic ECG.
\begin{figure}
\centerline{\psfig{file=rsrr.eps,width=7.75cm}}
\end{figure}
As a consequence of constructing the model with a variable angular frequency $\omega(t)$, the time taken to move from the Q event to the T event, known as the QT-interval, varies with the RR-interval on a beat-to-beat basis. The relationship between the QT-interval and the RR-interval is linear as shown in Fig. 10. Such a linear relationship has been reported for real ECGs and has been used to calculate a corrected QT-interval [4]. It is interesting that this relationship is a direct consequence of the model. Furthermore it may be possible to use the model to assess how much of the variation in the QT-interval is due to RR-interval variability so that this effect can be factored out.
Figure 10: QT-intervals versus RR-intervals for the synthetic ECG.
\begin{figure}
\centerline{\psfig{file=qtrr.eps,width=7.75cm}}
\end{figure}