This paper describes how to evaluate an automated ECG analyzer using
available annotated ECG databases and software, in compliance with
standard evaluation protocols. These protocols have been adopted as
parts of the American National Standard for Ambulatory
Electrocardiographs (ANSI/AAMI EC38:1998, and its predecessor,
ANSI/AAMI EC38:1994), and the American National Standard for
Testing and Reporting Performance Results of Cardiac Rhythm and ST
Segment Measurement Algorithms (ANSI/AAMI EC57:1998). They include
earlier evaluation protocols developed for an AAMI Recommended
Practice, Testing and Reporting Performance Results of
Ventricular Arrhythmia Detection Algorithms (AAMI ECAR, 1987). It
will be most useful to readers who plan to use the suite of evaluation
software included in the WFDB Software Package (http://www.physionet.org/physiotools/wfdb.shtml); this
suite of software includes the reference implementations of the
evaluation protocols specified in EC38 and EC57.