.TH RXR 1 "15 June 1995" "MIT DB software 9.3" "DB applications" .SH NAME rxr \- ANSI/AAMI-standard run-by-run annotation comparator .SH SYNOPSIS \fBrxr -r \fIrecord\fB -a \fIreference-annotator test-annotator\fR [ \fIoptions ... \fR ] .SH DESCRIPTION .PP This program implements the run-by-run comparison algorithm described in \fITesting and Reporting Performance Results of Ventricular Arrhythmia Detection Algorithms\fR, an AAMI Recommended Practice (AAMI ECAR-1987). Using options \fB-C\fR, \fB-L\fR, or \fB-S\fR, \fIrxr\fR implements the run-by-run comparison algorithm specified by the current American National Standard for ambulatory ECG analyzers (ANSI/AAMI EC38-1994). \fIrxr\fR is the reference implementation of these algorithms, and must be used to obtain the run-by-run performance statistics cited in EC38 in order to be in compliance with the standard. The EC38 comparison algorithm includes all of the ECAR statistics (so that both sets of requirements are satisfied by using the EC38 algorithm), with additional statistics on detection of runs of supraventricular ectopic beats (SVE runs). .PP Input to this program consists of two annotation files associated with the same \fIrecord\fR. One of these is designated the \fIreference\fR annotation file, the other the \fItest\fR annotation file (called the `algorithm' annotation file in EC-38 and in ECAR). .PP \fIOptions\fR include: .TP \fB-c \fIfile\fR Append condensed reports (ECAR Table 10 format) to \fIfile\fR. .TP \fB-C \fIfile\fR As for \fB-c\fR, but report SVE run statistics also. .TP \fB-f \fItime\fR Begin the comparison at the specified \fItime\fR (default: 5 minutes after the beginning of the record). .TP \fB-h\fR Print a usage summary. .TP \fB-l \fIfile\fR Append line-format reports (ECAR Table 13 format) to \fIfile\fR (see below). .TP \fB-L \fIfile file2\fR As for \fB-l\fR, but report SVE run statistics in \fIfile2\fR. .TP \fB-s \fIfile\fR Append standard reports (ECAR Table 3 format) to \fIfile\fR. .TP \fB-S \fIfile\fR As for \fB-s\fR, but report SVE run statistics also. .TP \fB-t \fItime\fR Stop the comparison at the specified \fItime\fR (default: the end of the record if it is defined, the end of the reference annotation file otherwise; if \fItime\fR is 0, the comparison ends when the end of either annotation file is reached). .TP \fB-v\fR Verbose mode (list all discrepancies; see below). .TP \fB-w \fItime\fR Set the \fImatch window\fR (default: 0.15 seconds; see below). .PP .PP At most one of \fB-c\fR, \fB-C\fR, \fB-l\fR, \fB-L\fR, \fB-s\fR, and \fB-S\fR can be given as an option. If `\fB-\fR' is given as a \fIfile\fR argument, reports are written on the standard output. If no options are specified, \fIrxr\fR writes standard reports on the standard output (equivalent to using the option \fB-s -\fR). The output generated by selecting \fB-l\fR or \fB-L\fR includes column headings only if a \fIfile\fR other than `\fB-\fR' is specified, and only if the specified \fIfile\fR does not already exist. In this way, \fIrxr\fR can be used repeatedly to build up a line-format table for multiple records, for further processing by \fIsumstats\fR(1). .PP The \fB-v\fR option specifies that each mismatch is described on the standard output in a format similar to: .br 3/5(120188-121065) .br where the first number is the reference run length, the second is the test run length (each of these is between 0 and 6), and the numbers in parentheses indicate the location of the match window in sample intervals. .PP The shell variable \fBDB\fR should be set and exported (see \fIsetdb\fR(1)). .SH BUGS .PP Since \fIrxr\fR performs multiple passes over its input files, it cannot be used at the end of a pipe. .SH SEE ALSO bxb(1), ecgeval(1), epic(1), mxm(1), setdb(1), sumstats(1) .br \fIEvaluating ECG Analyzers\fR (in the \fIECG Database Applications Guide\fR) .br \fITesting and Reporting Performance Results of Ventricular Arrhythmia Detection Algorithms\fR (publication AAMI ECAR-1987) .br \fIAmerican National Standard for Ambulatory Electrocardiographs\fR (publication ANSI/AAMI EC38-1994) .br The last two publications are available from AAMI, 3330 Washington Blvd., Suite 400, Arlington, VA 22201 USA.