Next: Obtaining Reference Heart Rate Data
Up: Evaluating an ECG Analyzer
Previous: Presenting Signals to the Analyzer
For any ambulatory ECG monitor that incorporates automated analysis functions,
the EC-38 standard requires the manufacturer to implement and disclose a
method for producing test annotation files. Independent evaluators should seek
assistance from the manufacturer in any case, since the manufacturer's
interpretation of the device's outputs in the language of EC-38 is definitive
(in effect, the annotation file generation technique becomes part of
the system under test). Note that generation of annotation files need not be
synchronous with data acquisition; a device might conceivably store all of the
necessary data until the end of the test, and only then write the file.
Neither does the standard require that an annotation be determined within any
fixed amount of time, as would be expected of devices designed to trigger
pacing, for example. Furthermore, EC-38 specifically allows for the
possibility that the device under test might not produce the annotation file
directly. If any external hardware or software is required to do so, however,
it must be made generally available or specified in sufficient detail by the
manufacturer to permit an independent evaluator to obtain test annotation
files.
Annotation files contain a label (an annotation) for each beat and for certain
other features of the signals, such as rhythm and ST changes. Annotations are
stored in time order in annotation files. The ``time'' of an annotation is
that of the sample in the signal file with which the annotation is
associated. The DB library (included in the DB software package)
includes C-callable functions (getann and putann) for reading and
writing annotations. In a C program, annotations appear as data structures
containing a 32-bit time field together with a pair of 8-bit fields that
encode the annotation type and sub-type (anntyp and subtyp [sic],
respectively), and a variable-length aux field usually used to store
text. In annotation files, these annotation structures are usually stored in
a variable-length bit-packed format averaging slightly more than 16 bits per
annotation.
Test annotation files may include the following:
- Beat annotations. These need not coincide precisely with the
reference beat annotations, since the evaluation protocol allows a time
difference of up to 150 ms between each pair of matching beat annotations.
All beat annotations are mapped during the evaluation process into the set
{ N, V, F, S, Q } (corresponding to normal, ventricular ectopic, ventricular
fusion, supraventricular ectopic, and unclassifiable or paced beats
respectively); devices need not be capable of producing all of these
annotations, but any beat annotations that they do produce will be translated
into one of these types. The standard specifies the mapping used for the
anntyp values defined in <ecg/ecgcodes.h>. (This file is
supplied with all of the existing CD-ROM databases.) Any beat annotations that
appear in the first five minutes of a record (the ``learning period'') are
ignored in the evaluation process. The remainder of the record (the ``test
period'') must be fully annotated. Note in particular that the last beat of
some records may be very close to the last sample; since the analyzer may
reach the end of the input signals before producing an annotation for the last
beat, it may be necessary to ``pad'' the input data for a few seconds at the
end of the record to permit the analyzer to emit its final beat annotation.
- Shutdown annotations. If the device suspends its analysis
because of poor signal quality or for any other reason, it should mark the
periods during which analysis is suspended. The evaluation software tallies
beats missed during such periods separately from beats missed at other times.
The beginning of each period of shutdown may be marked using a NOISE
annotation with , and the end of each period of shutdown
may be marked using a NOISE annotation with (see
the source for bxb for notes on other acceptable methods of marking
shutdown).
- Ventricular fibrillation annotations. The beginning and end
of each detected episode of ventricular fibrillation should be marked using
VFON and VFOFF annotations.
- Other rhythm annotations. These should include
RHYTHM annotations marking the beginning and end of each detected episode
of atrial fibrillation. The beginning of each episode should be marked with an
``(AFIB'' rhythm annotation, i.e., an annotation with anntyp =
RHYTHM and aux =
"\05(AFIB"
, where ``\05
'' is C notation
for a byte with the value 5 (ASCII control-E). Non-empty aux fields
always begin with a byte that specifies the number of data bytes that follow;
in this case, the five characters (( A F I B) of the string. The end of
each episode should be marked with any other rhythm annotation (for example,
``(N''). - Heart rate measurements. Each type of heart rate measurement
(including any heart rate or RR interval variability measurements) made by the
device under test should be assigned a measurement number, m, between 0 and
127. A MEASURE annotation should be recorded for each heart rate
measurement, with and with the measurement in the aux
field, as an ASCII-coded decimal number.
- ST deviation measurements. If available, these should be
provided in the aux fields of beat annotations, as ASCII-coded decimal
numbers indicating the deviations in microvolts from reference levels
established for each signal from the first 30 seconds of each record. For
example, ``25 -104'' indicates a 25 V elevation in signal 0 and a 104
V depression in signal 1. If ST measurements are omitted from any beat
annotation, the evaluation software assumes they are unchanged from their
previous values.
- Ischemic ST change annotations. These STCH annotations
should mark the beginning and end of each detected episode of ischemic ST
change. ST change annotations have additional information in the aux
field as for rhythm annotations: the beginning of each episode is marked by an
``(STns'' annotation, and the end of each episode by a
``STns)'' annotation, where n indicates the signal affected
(``0'' or ``1''), and s indicates ST elevation (``+'')
or depression (``-''). n may be omitted if the episode detection
criteria depend on features of both signals. The extremum of each episode may
optionally be marked with an ``ASTnsm'' annotation, where n
and s are defined as above, and m is the ST deviation in
microvolts, relative to a reference level established as above.
Note that only beat annotations are absolutely required in test annotation
files. ST deviation measurements within beat annotations, and the other
types of annotations listed above, only need to be recorded for devices
that are claimed by their manufacturers to provide optional features for
detection of ventricular or atrial fibrillation, measurement of ST deviations,
or detection of ischemic ST changes.
If the time units in the test annotation files are not the same as those in
the reference annotation files (for example, because xform was used to
change the sampling frequency of the signal files in a digital-domain test),
the time units must be rescaled before proceeding with the comparison. This
may be done by using xform to rewrite the test annotation files with
the original sampling frequency.
Details of the ST deviation measurement and episode detection criteria used in
producing the reference annotation files for the ESC database may be found in
several sources. Note, however, that many techniques
for measuring ST deviation and for detecting transient ischemic ST changes are
possible, and that to date the best evaluation results have been obtained for
analyzers using criteria that do not attempt to mimic those used by the human
experts who annotated the database.
Next: Obtaining Reference Heart Rate Data
Up: Evaluating an ECG Analyzer
Previous: Presenting Signals to the Analyzer
George B. Moody (george@hstbme.mit.edu)
Sat May 24 04:20:05 EDT 1997